A modern fume cupboard relies on more than its enclosure, sash and exhaust system. The controls behind the equipment also have an important role in managing how the cupboard starts, operates, responds to abnormal conditions and shuts down.
The Dynasafe Control System MK5 gives laboratory users a structured way to operate compatible Dynaflow fume cupboards while providing clear information about the cupboard's current status. Its functions cover airflow-related monitoring, operating sequences, alarms, timers and emergency controls, all through a panel designed around day-to-day laboratory use.
For laboratories considering new equipment or reviewing an existing installation, it is useful to understand how the control system fits into Dynaflow's wider range of fume cupboards and what its functions mean in practice.
The Dynasafe MK5 control system is the fifth-generation Dynasafe fume cupboard control platform used with compatible Dynaflow equipment. It may also be referred to in Dynaflow documentation as the MK5, Mark 5 or P-5 control panel.
At the user level, the Dynasafe MK5 control panel provides information about the cupboard's operating mode and status while giving the operator access to the main controls required during routine use. These include controls for starting and stopping the cupboard, lighting, compatible airflow boost functions and alarm muting.
Behind the panel, the controller manages the sequence between the cupboard, its ventilation system and associated services. This allows the fume cupboard control system to check operating conditions before services become available and to respond when monitored conditions move outside the required range.
The MK5 forms part of a broader range of Dynaflow products for laboratory fume management rather than functioning as a standalone safety device.
For everyday users, the value of the Dynaflow MK5 lies in the amount of operating information and control available from one interface.
Clear operating status. The display shows information such as the current operating mode, time and system status. Users can see whether the cupboard is off, purging or ready for normal operation instead of relying solely on whether the fan appears to be running.
Structured start-up and shutdown. The controller manages defined sequences around normal use. These sequences give the ventilation system time to establish operation before laboratory services are enabled and allow continued extraction after work has finished.
Monitoring and alarms. The controller monitors a number of inputs associated with cupboard operation. Depending on the installation and configuration, abnormal conditions can generate visible and audible warnings.
Process control. Timer functions allow operating periods to be scheduled when appropriate for the laboratory's workflow.
Emergency isolation. An emergency isolator on the panel can shut down the cupboard and isolate electrical and gas services where required.
Compatible airflow functions. When a cupboard is fitted with Dynaflow's Syncroflow® system, the Boost control can increase airflow extraction. Its operation depends on the equipment and configuration installed with the cupboard.
These functions give the fume cupboard controller a more active role in the operating sequence while still keeping routine controls accessible to users.
One of the most important things for users to understand is that pressing the FAN button does more than switch a fan on or off. The system moves through four main operating modes: OFF, PRE-PURGE, RUNNING and POST-PURGE.
In OFF, the cupboard is in its normal inactive state when chemical handling is not taking place. Electrical and gas services for ancillary equipment within the cupboard are unavailable.
When the operator presses FAN to begin work, the system enters a 60-second fume cupboard pre-purge. During this period, airflow is established and relevant equipment is given time to start and stabilise. Before the controller proceeds, it checks that monitored inputs are within normal operating conditions.
If the required conditions are satisfied, the controller enters RUNNING mode. Electrical and gas services then become available, and the display confirms that the cupboard has entered normal operation.
When the user presses FAN at the end of the handling process, the controller starts a 20-minute fume cupboard post-purge. Services are switched off while the extraction fan continues to operate. Once that period is complete, the cupboard returns to OFF.
Understanding these stages helps laboratory staff recognise why there can be a deliberate delay before services become available and why extraction continues after work has finished.
The MK5 monitors several inputs so that users can be alerted when the system detects an abnormal operating condition. The fume cupboard alarm system can provide both an audible warning and information on the panel to help identify what requires attention.
For example, Dynaflow's documentation identifies low airflow as an alarm condition. When airflow in the exhaust duct falls below the configured threshold, electrical and gas services are isolated until the problem has been corrected. Other monitored conditions can include mains power failure, backup battery status and sash position, depending on the configuration.
The controller also performs checks during the pre-purge stage. If an alarm condition is present, the cupboard will not simply proceed into normal operation. During RUNNING mode, an alarm can cause services to be switched off while the system signals the problem.
The emergency isolator provides another level of user control by allowing electrical and gas services to be isolated during an emergency.
An alarm should always be treated as a condition that requires attention. Muting an audible alarm does not correct its cause, and safety-related controls should never be bypassed. Users should follow their laboratory procedures and arrange investigation by trained personnel where required.
The MK5 also incorporates fume cupboard timer functions that can support laboratories with planned or recurring operating periods.
Dynaflow documentation states that up to four start-stop sequences can be programmed across a seven-day period. Each sequence can be set for a nominated day and time, allowing the cupboard to begin its operating sequence automatically.
At the scheduled start time, the system still performs its 60-second pre-purge rather than moving directly into RUNNING mode. When the programmed operating period ends, it moves through the normal 20-minute post-purge before returning to OFF.
This can be useful where laboratory activities need planned ventilation periods, provided scheduling is appropriate for the process and consistent with site procedures.
Configuration and service-level settings should be left to personnel authorised to work with the equipment. Dynaflow documentation notes that password-protected configuration functions are intended primarily for commissioning, maintenance and troubleshooting.
From the operator's perspective, the MK5 brings several practical functions together through one fume cupboard control panel:
The MK5 contributes to the overall fume cupboard safety system, but the controller alone cannot make a laboratory process safe. Effective containment also depends on the cupboard design, correct use, suitable ventilation, maintenance, testing, appropriate sash positioning and the laboratory's own risk controls.
The controller works as one component within a complete fume management system. Effective fume cupboard ventilation control also depends on the exhaust arrangement, fan performance, ductwork, sash position and the physical design of the cupboard itself.
Dynaflow lists the MK5 across multiple fume cupboard solutions. The Dynaflow GRP Fume Cupboard, for example, includes the Dynasafe MK-5 panel within its product specifications, while specialised equipment such as the Dynaflow PER Fume Cupboard also incorporates MK5 control functionality.
Safe operation continues beyond the controls. Users need to work with the sash in the appropriate position, avoid obstructing airflow and follow the procedures established for their laboratory and application.
Regular testing and maintenance are also part of that broader approach. Dynaflow provides fume cupboard testing services and uses custom testing software intended to produce consistent, calibrated and auditable results.
Applicable Australian Standards, site procedures, risk assessments and manufacturer guidance should continue to determine how each installation is operated and maintained.
Older fume cupboards can remain mechanically serviceable while their controls become outdated, difficult to support or less suited to the way the laboratory now operates.
Upgrading to a Dynasafe fume cupboard controller may provide access to improved operating controls, monitoring, purge sequencing and other functions available within the MK5 platform. Compatibility, however, needs to be assessed for the individual installation.
The condition and design of the existing cupboard, electrical systems, ventilation arrangement, connected services and any specialised equipment all need to be considered before an upgrade is recommended. Dynaflow states that it has experience retrofitting its control systems to existing fume cupboards, but the scope of work depends on the equipment involved.
Any control upgrade should be assessed and carried out by appropriately qualified personnel. This helps ensure the changes suit the existing equipment and do not interfere with the cupboard's established safety functions or site requirements.
The Dynasafe® MK5 gives laboratory users a structured operating sequence, clearer system feedback and access to functions such as monitoring, alarms, process timers, emergency isolation and controlled purge cycles.
Used as part of a correctly designed, maintained and tested fume cupboard installation, these functions can make day-to-day operation easier to understand while giving users more visibility of the equipment's operating condition.
For information about MK5 controls, compatibility assessments, upgrades, fume cupboard testing or maintenance, contact Dynaflow to discuss the requirements of your laboratory.
Contact Dynaflow
 
 
When FAN is pressed to start the cupboard, the controller enters PRE-PURGE mode for 60 seconds. During this period, airflow is established and relevant equipment is allowed to stabilise. The system checks its monitored inputs before allowing the cupboard to enter RUNNING mode and enabling electrical and gas services.
Post-purge is the controlled extraction period that occurs after normal operation has been stopped. On the MK5, services are switched off while the extraction fan continues to operate for 20 minutes. The cupboard then returns to OFF mode.
Check the information displayed on the panel and follow your laboratory's established procedures. Do not bypass the alarm or safety controls. Where the cause cannot be safely identified and corrected within normal user procedures, contact appropriately trained service personnel. Dynaflow's broader FAQ resources also provide information relevant to laboratory equipment and controls.
An upgrade may be possible, but the existing installation should be assessed first. Compatibility can depend on the cupboard, ventilation system, electrical arrangement, services and existing control equipment. Dynaflow can assess the installation and advise on suitable upgrade options rather than assuming that every existing cupboard can accept the same retrofit.